Loading Slide Racks; Loading Slides - Siemens ADVIA 2120 Operator's Manual

Table of Contents

Advertisement

5. Select
6. When prompted to prime lines, select
select
7. If bubbles remain in the reagent lines, prime the lines again using the Reagent
Inventory window.

Loading Slide Racks

Load at least two racks into the
rack loader. The rack loader can
hold up to 9 racks at a time.
Load the racks as shown below,
with the slide grooves facing
upwards.
You can load racks at any time.

Loading Slides

When the autoslide startup cycle runs, the slide dispenser must contain slides and there
must be racks in the rack loader.
Do not try to empty the slide dispenser. Finger injuries and sensor damage can occur if
you attempt to remove slides from the slide dispenser.
Microscope slides - storage and handling
When storing and handling microscope slides, follow these precautions:
Slides should be allowed to come to Lab temperature, before they are opened.
10-12
.
OK
.
OK
WARNING
Microscope slides and cover glasses should be rotated when stored.
To prevent moisture contamination, store the slides off the floor.
To minimize temperature and humidity changes, keep slides as far away as
possible from doors and heating or air conditioning ducts.
Remember, slides do not have to be old to be affected by moisture.
. The system allows only one prime cycles;
Yes
ADVIA Autoslide Slide Maker Stainer