Emptying The Stain Waste Container; Running The Daily Startup Manually - Siemens ADVIA 2120 Operator's Manual

Table of Contents

Advertisement

Use slides provided by Siemens
approved vendors only.
Size:
Description:
rounded corners, super frost
Avoid touching slide surfaces with
your fingers.
Control-painted areas of the slides
are all facing up.
Make sure slides are not stuck together before loading.
Make sure to load the slides in the correct orientation. The painted area must
be on the top of the slides and on the left-hand side of the stack.
Insert a stack of slides as shown above. The stack of slides must reach the
stack level detectors. The slide dispenser can hold up to 160 slides.

Emptying the Stain Waste Container

The stain waste is collected in a container located behind the system. Empty the container
periodically or whenever the system indicates that the container is full.
Stain waste must be collected using the tube and container specified by local regulation.
Stain waste contains methanol. Methanol is a solvent that may destroy an inappropriate
connection and cause leaks that are dangerous for the user and damaging to the
instrument.
Dispose of waste according to local regulatory requirements.
Make sure that the Autoslide module is not processing slides while you disconnect the
stain waste container
IMPORTANT
Local laws and regulations protect the environment and encourage resource conservation
by regulating the disposal of hazardous wastes. Because some of the wastes generated by
the analyzer may be classified as hazardous waste, you must be familiar with the
applicable hazardous waste handling and disposal laws and regulations in your area.

Running the Daily Startup Manually

Your system can be set to run the startup procedure at a preset time. However, if the
system is not set to do so, or if you are starting it at a different time, perform the
following steps.
ADVIA Autoslide Slide Maker Stainer
76 x 26 x 1mm
ground edges,
WARNING
10-13

Advertisement

Table of Contents
loading

This manual is also suitable for:

Advia 2120i

Table of Contents