Performance Characteristics: Correlation Data - Siemens ADVIA 2120 Operator's Manual

Table of Contents

Advertisement

Performance Characteristics: Correlation Data

The performance of the ADVIA 120 Hematology System was compared with the
performance of the Technicon H•3 RTC System (x) using whole blood from 60
normal donors and 70 hospital donors with each sample run in duplicate (n =
260).
The following regression statistics were computed using a protocol similar to that
recommended in the NCCLS document EP9-A, Method Comparison and Bias
Estimation Using Patient Samples; Approved Guideline (1995).
Parameter
%NEUT
%LYMPH
%MONO
%EOS
%BASO
%LUC
x = Technicon H•3 RTC system
y = ADVIA 120 Hematology system
Regulatory Information
Comparative System
or Method
Technicon H•3 RTC
Technicon H•3 RTC
Technicon H•3 RTC
Technicon H•3 RTC
Technicon H•3 RTC
Technicon H•3 RTC
Regression
r
Equation (y=)
1.02x - 0.6
0.997
1.00x + 0.8
0.997
0.85x - 0.3
0.943
0.87x + 0.2
0.979
0.67x + 0.0
0.772
0.92x + 0.6
0.944
Sy.x
1.6
1.7
0.8
0.4
0.2
0.7
9-53

Advertisement

Table of Contents
loading

This manual is also suitable for:

Advia 2120i

Table of Contents