Replacing The Perox Flowcell - Siemens ADVIA 2120 Operator's Manual

Table of Contents

Advertisement

2. Unscrew the securing cap to which the
power cable (
the lamp assembly out of the optics lamp
housing. The lamp assembly includes the
lamp, the power cable, and the securing
cap. Discard the entire assembly.
3. Insert the replacement lamp assembly.
Do not touch the glass bulb. If necessary, clean with lens paper and methyl
alcohol.
4. Replace the securing cap, then connect the red power connector.
5. Turn the analyzer on and align the lamp to maximize its output.
(See page 5-4 for instructions.)

Replacing the perox flowcell

Time: Installation - 15 minutes
Analyzer mode: Standby
All products or objects that come in contact with human or animal body fluids
should be handled, before and after cleaning, as if capable of transmitting
infectious diseases. Wear facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended for potentially
infectious specimens in Protection of Laboratory Workers from Infectious
Disease Transmitted by Blood, Body Fluids, and Tissue, 2nd edition; Approved
Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user
protection and it can be used as reference material for instructions on laboratory
safety.
IMPORTANT
An inspection scope must be available for this procedure. Do not remove the
flowcells unless you are trained in how to align and focus them.
To remove the flowcell
1. Make sure that the analyzer is in the Standby mode.
Troubleshooting the Analyzer
) is attached. Carefully pull
2
CAUTION
Checkout - 15 minutes
BIOHAZARD
VIEW FROM
BACK OF PEROX
OPTICS ASSEMBLY
5-39