Ending Each Shift

Washing the System

Perform the system wash procedure at the end of each shift or work period (a
maximum of eight hours). After the laboratory shift with the largest number of
samples, run three system wash cycles; after other shifts, you need to run only
one wash cycle. In addition, if the number of samples in a shift exceeds 400,
perform one system wash after the 400th sample.
1. At the Utilities menu, select
2. Select
Start

Performing the End of Day Procedure

To reset SIDs, and close out the QC and Moving average statistics (Tools
View)
1. At the Customize menu, select
2. When the tools list displays, double-select
3. Select
selections from a previous End of Day operation now appear.
4. Select the
5. If desired, select the
into the Cumulative QC file.
Use the list box to select close out for all controls or a specific control.
6. If desired, select the
data point from moving average control statistics with a date earlier than the
date in the Up to box.
7. Select
To also backup and purge the data files (Tools Modify)
1. At Customize menu, select
2. When the tools list displays, double-select
3. Select the
4. If desired, select the
into the Cumulative QC file.
Use the list box to select close out for all controls or a specific control.
5. If desired, select the
data point from moving average control statistics with a date earlier than the
date in the Up to box.
3-10
System Wash
, select
.
Yes
to confirm that you have finished the workload. If saved, the
check box to perform a SID RESET.
SID RESET
Q.C. Close Out
MOV. AVG Close Out
to confirm and start.
OK
SID RESET
check box to perform a SID RESET.
Q.C. Close Out
MOV. AVG Close Out
Hydraulic Functions
1
for the
Number of Cycles
Tools View
.
End Of Day
check box to transfer the daily QC file
check box to obtain a cumulative
, and then select
System Setup
End Of Day
check box to transfer the daily QC file
check box to obtain a cumulative
.
, and then select
.
Tools Modify
.
Daily Routine
.