Adjustments

Collimator Slot Diaphragm

Prerequisites
Image rotation has been adjusted correctly.
Use the slot diaphragm keys at the control console for the adjustment.
Image reversal functions are disabled.
Use 2.1 mm Cu as prefilter.
Adjustment
1. Register an emergency patient to allow radiation release
2. Click on <Go>.
0 Degree Position
X
3. Release fluoro and move the slot diaphragm to the vertical position ( ↑ ) on the monitor.
4. Stop fluoro and click on <Go>.
I.I. Full Format Position
X
5. Release fluoro and open the slot diaphragm so that the blades begin to disappear at the
edge of the image.
NOTE
6. Stop fluoro and click on <Go>.
I.I. Zoom Format Position
X
7. Release fluoro and open the slot diaphragm so that the blades begin to disappear at the
edge of the image.
8. Stop fluoro and click on <Go>.
Init Off & Store Position
Siemens AG
Medical Solutions
Status bar: Init (LimitsOff/Autosearch) --> running --> success
Status bar: 0 Degree Position --> running --> success
For units with air kerma display it is important for software calcula-
tion that the slot diaphragm in "I.I. - Full Format Position" is
adjusted accurately to the edge of the visible I.I. field.
Status bar: I.I.-Full Format Position --> running --> success
Status bar: I.I.-Zoom Format Position --> running --> success