Functional Test Of C-Arm Lift Movement And Parking Brakes - Siemens ARCADIS Varic Start-Up

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

22

Functional test of C-arm lift movement and parking brakes

Fig. 4:
Switch on the main system (C-arm).
To move the C-arm downward, press the downward button.
The lifting column moves to position 1
ly. A stop signal (3 beeps) sounds simultaneously.
To lower the lifting column further, press the downward button once again.
The lifting column can then be moved by a further 5 cm to the lowest point, position 2
(2/Fig. 4 / p.
downward button is pressed.
To move the C-arm upward, press the upward button. No signal sounds in this case.
The lift motor remains stationary if both buttons are pressed simultaneously.
Loosen the different parking brakes and apply them again. Secure blocking of the rele-
vant movement must be assured.
Document the function of the C-arm movements, brakes and warning signals in the test
protocol, "C-arm movement and emergency stop"
stop / p.
ARCADIS Varic
C-arm movement functional test
22). In this area a signal (3 beeps) sounds for safety reasons each time the
40).
SPR2-310.815.01.05.02
02.05
General Start-up
(1/Fig. 4 / p. 22)
and remains there automatical-
(C-arm movement and emergency
Page 22 of 48