Measurements With The Oscilloscope; Swivel-Arm Rotation; Service Software - Siemens MAMMOMAT Novation DR Installation Manual

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Important Notes
Apply the protective strips after removing the covers from the stand. Remove the protec-
tive strips only when the covers are to be mounted or when vertical adjustment of the
swivel arm system is necessary.
Fig. 1 Protective strips for the metal curtain
Holders (2/Fig. 1) for storing the protective strips when not in use are provided on both
sides of the curtain.

Measurements with the oscilloscope

WARNING

Swivel-arm rotation

WARNING

Service Software

The MAMMOMAT Novation
Acquisition Workstation
The Acquisition Workstation service software is web-based and can be started from the
Options application menu.
BRICK
The BRICK service software can be accessed via the Acquisition Workstation.
Siemens AG
Medical Solutions
1
2
The existing ground conductor in the mains cable must under no
circumstances be disconnected when operating the oscilloscope.
A life-threatening hazard of electric shock exists.
For measurements in which a resulting ground loop may falsify
the measuring result, use the differential amplifier (difference
measurement).
Never rotate the swivel arm with the detector wing in the parking
position. This will damage the main cable harness.
DR
has various service software packages:
SPB7-250.815.02
Rev. 02
Page 3 of 4
09.04
CS SD 24
2 - 3
2
2
2
DR
MAMMOMAT Novation
System

Advertisement

Table of Contents
loading

Table of Contents