Setting The Value Of The Output Signal; Edit Mode - Fluke 6003A Operator's Manual

Three phase power calibrator
Hide thumbs Also See for 6003A:
Table of Contents

Advertisement

The state of the Product can be changed with the buttons located on the front
panel in one of the subsequent ways:
1. Change of function by pushing one of the direct function buttons
After you push one of the buttons (, , , ) the Product switches to
the selected function and sets the last setup parameters for that function.
Whenever a function is changed, the Product always goes to the STANDBY
position and the output terminals are disconnected. The P, E, V, and I
functions have softkeys that are shown in the display.
2. Connection /disconnection of output terminals
After  is pushed, the output terminals of the Product are connected. Only
those channels (phases) that are activated are connected. To disconnect
terminals, push .
3. Access to the Main Setup menu
After  is pushed, the Product shows the basic level of the setup menu. In
this menu you can change instrument setting and calibration data. The
previous menu is recalled after the Exit softkey is pushed.

Setting the Value of the Output Signal

Edit Mode

Parameters of the output signal can be changed in edit mode. Only parameters
shown in blue can be changed. The display can be switched to edit mode in
these ways:
Push a numeric button
Push 
Push any of the cursor buttons (, , ,)
Push the rotary knob
In edit mode, the edited value is highlighted with a blue background. Push  to
change to the next "blue" parameters. Push the Exit softkey to get out of edit
mode.
Features and Basic Operation