Text Off Key - Fluke Autoranging CombiScope PM3370B Reference Manual

Hide thumbs Also See for Autoranging CombiScope PM3370B:
Table of Contents

Advertisement

BRIEF CHECKING PROCEDURE
TIME cursors check:
-
Using a softkey, select time cursors (//).
Press the READOUT softkey and select ∆T from the menu.
-
-
Press the RETURN softkey.
Turn the ∆ control so that the dotted line is approximately 2 divisions to the
-
right of the dashed line.
-
Turn the TRACK control so that the dashed line coincides exactly with the first
negative edge of the signal.
Turn the ∆ control so that the dotted line coincides exactly with the second
-
negative edge of the signal.
Check for a time cursor readout of approximately 500 µs in the top of the
-
display area.
-
Press the ANALOG key ('ANALOG MODE' is displayed briefly). The
oscilloscope is now in analog mode.
-
Verify that the cursors are on the first and second negative slope of the
waveform. If necessary readjust them using the TRACK and ∆ controls.
Check for a time cursor readout of approximately 500 µs in the top of the
-
display area.

3.7 TEXT OFF key:

-
Press the TEXT OFF key and verify that the CRT softkey text disappears.
-
Press the TEXT OFF key again and verify that only the signal, the cursors, and
the cursor readout value are displayed.
Press the TEXT OFF key to obtain full information again in the viewing area.