Error display

Description
If hardware faults, software errors or diagnostic alarms occur in the transmitter, the message
"Error" appears in the measured value display.
A ticker appears in the bottom line of the display indicating the type of error. This diagnostic
information is also available via HART communication.
Display
Figure 6-3
The following messages may appear as a ticker in the bottom line of the display.
Ticker
"HARDWARE FIRMWARE ALARM"
"DIAGNOSTIC ALARM"
"SENSOR BREAK"
SITRANS P DS III with HART
Operating Instructions, 06/2013, A5E00047092-09
Example of error message
Meaning
Contains hardware faults such as:
incorrect checksum
incorrect EEPROM data
Defective EEROM
RAM fault
ROM fault
inconsistent data
EEPROMs not initialized
Is always displayed if
an event configured by the user is to be signaled
with an alarm.
For example:
Limit reached
Event counter for limit values exceeded
Calibration time expired
Current saturation reached
the status of one of the device variables is "BAD".
Appears when there is a sensor break.