Blind Setting Of Zero Point And Limit Point; Zero Point Calibration (Position Correction) - Siemens SITRANS P DS III Operating Instructions Manual

Pressure transmitter with hart
Hide thumbs Also See for SITRANS P DS III:
Table of Contents

Advertisement

7.5

Blind setting of zero point and limit point

● The start of scale value and the full scale value can be set without creating a reference
● Both values can be selected as any point within the sensor limits.
● The maximum downscaling is 1:100, depending on the series and the measuring range.
● You can use this function to realize rising and falling characteristic curves.
7.6

Zero point calibration (position correction)

Description
A zero point calibration is used to correct a zero point error resulting from the installation
position.
Procedure
● Pressurize the device or evacuate it (at absolute pressure, < 0.1 ‰ of the measuring
● Perform the zero point calibration using SIMATIC PDM or the HART Communicator.
● If you do not have a vacuum, perform a trimming of the lower sensor calibration point at a
See also
Trimming the sensor (Page 130)
SITRANS P DS III with HART
Operating Instructions, 06/2013, A5E00047092-09
pressure.
span).
known reference pressure.
CAUTION
For absolute pressure transmitters, the zero point is at vacuum. A zero point calibration
with a pressurized transmitter will lead to faulty settings!
Note
The effective measuring range is reduced by the amount of the upstream pressure.
Example:
With an upstream pressure of 100 mbar, the effective measuring range of a 1-bar
transmitter is reduced to a point between 0 and 0.9 bar.
Operating functions through HART
7.5 Blind setting of zero point and limit point
123

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents