Siemens SITRANS P DS III Operating Instructions Manual page 197

Pressure transmitter with hart
Hide thumbs Also See for SITRANS P DS III:
Table of Contents

Advertisement

DS III construction with PMC connection
Standard
Minibolt
Electrical connection
Torque for cable gland nut made of
Degree of protection for Han and
M12 connectors
Han 8D is identical to Han 8U.
1)
Design for absolute pressure (from the differential pressure series), differential pressure and flow rate
Weight
Material
Wetted parts materials
Seal diaphragm
Pressure caps and locking screw Stainless steel, mat. no. 1.4408 to PN 160, mat. no. 1.4571/316Ti for PN 420,
O-ring
Non-wetted parts materials
Electronics housing
Pressure cap screws
Mounting bracket
Measuring cell filling
SITRANS P DS III with HART
Operating Instructions, 06/2013, A5E00047092-09
Flush mounted
1
/
''
1
2
PMC Standard design
Flush mounted
1''
PMC Minibolt design
Cable inlet using the following cable glands:
Pg 13.5
M20 x 1.5
½-14 NPT
Han 7D/Han 8D plug
1)
M12 connector
Plastic
2.5 Nm (1.8 ft lb)
IP65
Approx. 4.5 kg (9.9 lb) for aluminum enclosure
Stainless steel, mat. no. 1.4404/316L, Hastelloy C276, mat. no. 2.4819, Monel,
mat. no. 2.4360, tantalum or gold
Hastelloy C4, 2.4610 or Monel, mat. no. 2.4360
FPM (Viton) or optionally: PTFE, FEP, FEPM and NBR
Non-copper aluminum die casting GD-AlSi 12 or stainless steel precision
casting, mat. no. 1.4408
Standard: Polyester-based paint
Option: 2 coats: Coat 1: epoxy-based; coat 2: polyurethane
Stainless steel nameplate
Stainless steel
Steel or stainless steel
Silicone oil
Neobee M20
Inert liquid
(max. 120 bar g (2320 psi g) for oxygen measurement)
Metal
4.2 Nm (3.1 ft lb)
Technical data
11.6 Construction
Stainless steel
4.2 Nm (3.1 ft lb)