Calibration; Calibration Equipment; Table 6-1: Load Resistors And Current - Ametek 2003RP User And Programming Manual

Ac power source
Table of Contents

Advertisement

6

Calibration

Routine calibration should be performed every 12 months. Non-routine calibration is only
required if a related assembly is replaced or if the periodic calibration is unsuccessful.
Calibration can be performed through the front panel by putting the 2003RP in Calibration
mode (CAL mode) or over the RS232C or IEEE-488 bus if the interface option is installed.
For bus calibration, the PGUI32 program may be used. Refer to the PGUI32 help file
supplied with the program for instructions on calibration over the remote control bus.
All routine calibrations can be performed without removing the top cover (closed case
calibration). For the –AV option, program 115V or 230V for 135V or 270V respectively.

6.1 Calibration Equipment

Digital Multimeter (DMM):
100 milliohm Current Shunt:
Load Bank:
4
Some 2003RP-AV may be factory configured for single voltage range only. In this case, the 230 V
range is not available.
Model
Range
2003RP
135 V
270 V
2003RP-AV
115 V
230 V
2003RP-HV
156 V
312 V

Table 6-1: Load Resistors and Current

Two (2) Fluke 8506A or 8508A
Isotek Model RUG-Z-R100-0.1 or equivalent
Various power load resistors will be needed
Current
Load
max.
5.0 A
27
2.5 A
108
5.9 A
20
4
2.9 A
78
4.3 A
36
2.16 A
144

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents