Table 10-4: *Rst Default Parameter Values; Table 10-5: Status Registers - Power On Conditions - Ametek 2003RP User And Programming Manual

Ac power source
Table of Contents

Advertisement

10.10.6
*RST
This command resets the AC source to a setting defined by the values in the register defined
by PON if valid or by the following factory-defined states:
Item
VOLT
CURR
OUTP
OUTP:STARt:STATe
NOTE: The factory defaults cannot be used on units with the -AV option. Instead, one
of the setup registers 0 through 7 must be selected for the power on state.
Executing the *RST command will recall the settings from the assigned setup
register. Use the SYST:PON to select one of the available setup registers for
the power on and *RST states.
Command Syntax
Parameters
Related Commands
Register
- ESE register
- ESR register
- STB register
- SRE register
Note: Sending the *RST command will cause the unit to recall the REMOTE or LOCAL state
based on the setting of the SYSTem:KLOCk command. This may put the unit in local if this
state is set to 0. To avoid this, either set the SYSTem:KLOCk state to 1 or follow the *RST
command with a SYST:REM command. See SYSTem:KLOCk for information on the power
up remote / local state.

Table 10-4: *RST Default Parameter Values

Value
5.0 V
5.0 A
0 (OFF)
0 (OFF)
*RST
None
*SAV

Table 10-5: Status Registers - Power on Conditions

Status on Power On
All 0‟s
PON bit = 1, all others 0
All 0‟s
All 0‟s
Item
FREQ
VOLT:RANGe:LEVel
VOLT:RANGe:AUTO
DISP:MODE
OUTP:STARt:PHASe
Status after *RST
All 0‟s
PON = 0, all others unaffected
All 0‟s
All 0‟s
Value
60 Hz
135 V
0 (OFF)
1 (FREQ)
0