CALibrate:MEASure:POWer[:FSCale] <NRf>
This command affects the calibration of the power measurement at full scale.
Command Syntax
Parameters
Examples
Query Syntax
Returned Parameters
Related Commands
CALibrate:MEASure:POWer:ZERO <NRf>
This command affects the calibration of the power measurement zero offset.
Command Syntax
Parameters
Examples
Query Syntax
Returned Parameters
Related Commands
CALibrate:MEASure:VOLTage[:FSCale] <NRf>
This command affects the calibration of the rms voltage measurement at full scale.
Command Syntax
Parameters
Examples
Query Syntax
Returned Parameters
Related Commands
CALibrate:MEASure:VOLTage:ZERO <NRf>
This command affects the calibration of the rms voltage measurement zero offset.
Command Syntax
Parameters
Examples
Query Syntax
Returned Parameters
Related Commands
CALibrate:MEASure:POWer[:FSCale] <NRf>
<NRf> (value range -127 to +127)
CAL:MEAS:POW 78
CALibrate:MEASure:POWer?
<NR1> (value range -127 to +127)
CALibrate:MEASure:CURRent
CALibrate:MEASure:POWer:ZERO <NRf>
<NRf> (value range -127 to +127)
CAL:MEAS:POW:ZERO -5
CALibrate:MEASure:POWer:ZERO?
<NR1> (value range -127 to +127)
CALibrate:MEASure:POWer
CALibrate:MEASure:VOLTage[:FSCale] <NRf>
<NRf> (value range -127 to +127)
CAL:MEAS:VOLT 56
CALibrate:MEASure:VOLTage?
<NR1> (value range -127 to +127)
CALibrate:MEASure:CURRent
CALibrate:MEASure:VOLTage:ZERO <NRf>
<NRf> (value range -127 to +127)
CAL:MEAS:VOLT:ZERO +4
CALibrate:MEASure:VOLTage:ZERO?
<NR1> (value range -127 to +127)
CALibrate:MEASure:VOLTage