Figure 3-6: Input And Output Wiring; Table 3-1: Maximum Output Wiring Lengths - Ametek 2003RP User And Programming Manual

Ac power source
Table of Contents

Advertisement

Output Current
Wire Size
5.0 A
0.75 mm
RMS
5.0 A
AWG 18
RMS
5.0 A
1.0 mm
RMS
5.0 A
AWG 16
RMS
5.0 A
1.5 mm
RMS
5.0 A
AWG 14
RMS
The Remote Sense inputs must be connected or an output voltage 7% higher than the
programmed output will be generated. If the 2003RP is configured for constant voltage, a
fault will be generated.
AC
Output
Ferrite
Clamp
Maximum wire length from output to load, 1% drop
2
2
2

Table 3-1: Maximum Output Wiring Lengths

AC Output
Safety Cover
HI LO
GND
AC
Input

Figure 3-6: Input and Output Wiring

6 m
20 ft
6.4 m
21 ft
8 m
26 ft
8.5 m
28 ft
12 m
39 ft
14 m
45 ft
AC Input
Safety Cover
L
N
Ferrite
L N Gnd

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents