7.
After the paper sensor 1 detects that the calibration plate passed, moves for a specified length, the
calibration plate stops at the standby position.
• Paper sensor 1: DARK to LIGHT
• Calibration plate advance motor: ON (reverse)/OFF
If paper sensor 1 does not turn LIGHT
•
No. 06554-00006
Calibration Plate advance error.
8.
Apply pressure to the measuring position without the calibration plate.
• Pressure solenoid: OFF
9.
Judge and register the result of the calibration plate, calibrate the colorimeter.
For details of the judgment, refer to
56600
Judgment operation of setup print.
56600
Colorimeter unit operation
3/3