No. 05916[F] Solution Level Of The Processing Tank Goes Below The Processable Level. P; No. 05917[F] Solution Level Of The Processing Tank Goes Below The Processable Level. P; No. 05918[F] Solution Level Of The Processing Tank Goes Below The Processable Level. Ps; No. 05919[F] Solution Level Of The Processing Tank Goes Below The Processable Level. Ps - Frontier LP7500 Service Manual

Qss-37 series
Table of Contents

Advertisement

Distributed by: minilablaser.com
No. 05916[F]
Solution level of the processing tank goes below the processable level. P1
No. 05917[F]
Solution level of the processing tank goes below the processable level. P2
No. 05918[F]
Solution level of the processing tank goes below the processable level. PS1
No. 05919[F]
Solution level of the processing tank goes below the processable level. PS2
No. 05920[F]
Solution level of the processing tank goes below the processable level. PS3
No. 05921[F]
Solution level of the processing tank goes below the processable level. PS4
Error message release
YES
Condition
Suffix number
• # # #When the processable level decreases and addition of specific amount water is carried, # # #the
00000
• When the product starts up with program timer and filling up the moisture which evaporated is carried out,
• If Pump Output Amount Measurement of Pump Output Amount Setting and Measure Auto Cleaning Water Output
Amount of Auto Cleaning Water Output Amount Setting are not performed at system installation, the above errors may occur.
• The causes are shown below.
1. Even if specified amount of water is supplied, # # # processable level does not turn on.
2. Even if each replenishment pump operates for a specified time, specified amount of solution is not replenished.
N
OTE
• If the processing solution level detector of the processing solution float switch turns off, all the processing solution heaters and
circulation pumps turn off.
Countermeasure
1. Perform
Taking countermeasures after the errors No.05916 to No.05921 occurs.
4630
Check Point
Check if the processing solution is leaking from the tank.
Check if the float of the processing solution float switch moves up and down smoothly.
Check the status of the float switches of each processing solution via Input Check.
Check each replenishment pump output amount again.
Check the output amounts of auto cleaning pump + each automated cleaning valve again.
Select Operator Selections→Processor, and change the Evaporation Correction setting from OFF to
*1
ON.
*1. If Evaporation Correction is set to OFF, water is not replenished automatically although # # # processable level decreases.
Diagnosis
Processing solution float switch
processable level detector keeps off.
# # # the processable level detector keeps off.
When the level is incorrectly detected
45900
Condition
45900
Error message: F replenishment
-
-
35220
33003
33004
-
Manual No.
63280
12/16