Square Wave Output (For U1252B) - Agilent Technologies U1251B User's And Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1251B:
Table of Contents

Advertisement

Square Wave Output (for U1252B)

0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150, 200, 240, 300, 400, 480, 600, 800, 1200, 1600,
2400, 4800
N O T E
Agilent U1251B/U1252B User's and Service Guide
The square wave output function can be used to generate a
PWM (pulse width modulation) output or provide a
synchronous clock source (baud rate generator). You can
also use this function to check and calibrate flow- meter
displays, counters, tachometers, oscilloscopes, frequency
converters, frequency transmitters, and other frequency
input devices.
1 Turn the rotary switch to
display setting is 600 Hz on the secondary display and
50% duty cycle on the primary display.
2 Press
or
to scroll through the available frequencies
(choose from the available 28 frequencies):
Frequency (Hz)
Pressing
is the same as pressing
3 Press
to select the duty cycle (%) on the primary
display.
4 Press
or
to adjust the duty cycle. Duty cycle can be
set for 256 steps and each step is 0.390625%. The display
only indicates the best resolution with 0.001%.
Features and Functions
position. The default
.
3
69