Performance Verification Tests; Table 1-2 Performance Verification Tests - Agilent Technologies U1231A Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1231A:
Table of Contents

Advertisement

1
Calibration Procedures

Performance Verification Tests

Performance Verification Tests

N O T E

Table 1-2 Performance verification tests

Step Test function
1
Z
Turn the rotary switch
LOW
to the
position.
8
Use the performance verification tests to verify the
measurement performance of the instrument. The
performance verification tests use the instrument's
specifications listed in the U1231A/U1232A/U1233A User's
Guide (available for download at
www.agilent.com/find/hhTechLib).
The performance verification tests are recommended as
acceptance tests when you first receive the instrument. The
acceptance test results should be compared against the one
year test limits. After acceptance, you should repeat the
performance verification tests at every calibration interval.
If the multimeter fails the performance verification tests,
adjustment or repair is required.
Ensure that you have read the
running the performance verification tests.
Range
5520A input
600 V
"Test Considerations"
Error from nominal 1 year
U1231A
30 V
±0.9 V
U1231A/U1232A/U1233A Service Guide
on page 7 before
U1232A
U1233A
±0.9 V
±0.9 V