Defining Limit Line Tables - Agilent Technologies E5071C Compliance Tests

Mhl (mobile high-definition link) specification version 2.0; agilent method of implementation (moi) for mhl cables; ena option tdr
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Agilent MOI for MHL Cables Compliance Tests
4. Press
Display
-1/(2*PI*xAxis*im(50*(1+S11)/(1-S11)))".
5. Check Enabled to enable the equation on trace.
6. Click Apply.
7. Click Close.
8. Press
Scale
> Set Scale/Div to 100 pU/div.
9. Press
Scale
> Set Reference Value to -300 pU.

6.14. Defining Limit Line Tables

1. Press
Trace Next
2. Press
Analysis
> Limit Line and turn it ON to display limit lines.
3. Press
Analysis
> Edit Limit Line to edit the limit line table.
4. Press
Analysis
> Limit Test and turn it ON.
5. Press
Analysis
> Limit Test > Fail Sign to switch the fail sign ON/OFF. When turned
on, the Fail sign is displayed on the E5071C's screen, if one or more failed traces are
within the channel.
6. Press
System
> Misc Setup > Beeper > Beep Warning to turn ON/OFF the warning
beeper.
>
Equation
Editor...
to select trace to set the limit line table.
>
Enter
35
"Cs
an
equation
=