Optional Settings: Administrator Options - h/p/cosmos 150/50 Operating And Service Manual

Table of Contents

Advertisement

10 Optional settings: administrator options

The administrator options are only accessible through h/p/cosmos staff and authorised service engineers. Access is only
possible with a special code which is not given here.
Changing an administrator option might cause serious problems if you are not familiar with all details. Therefore the
h/p/cosmos service has to be contacted before modifying administrator options.
For optional settings on devices without UserTerminal, an external UserTerminal or a connected PC with the software
h/p/cosmos para control
List of administrator options
Option
Description
OP01
Reset of error messages E20, E21,
E30, E31, E32, E41, E50 and E51
OP02
Selection of the OEM code
OP03
Selection of device type
OP04
Maximum speed for reverse belt
rotation
(This option is only available for
treadmills, not for ladder
ergometers).
© 2013 h/p/cosmos sports & medical gmbh
is required.
®
Comment / Display
The required maintenance work has to be performed before deleting
error messages. Confirmation in the display
This option only resets the error message. If the error still exists, the
error message will appear again.
By selecting the OEM code, the functions and settings of the
UserTerminal are adapted to the requirements of the listed OEM
partner.
The following OEM codes can be selected:
0 = h/p/cosmos (standard)
1 = JAEGER / VIASYS / CARDINAL HEALTH / CARE FUSION
2 = Ergo-Fit (1997-98)
3 = Proxomed / Kardiomed
4 = KISTLER Gaitway
6 = COSMED
9 = SCHILLER
Not all companies listed above are or were OEM partners of
h/p/cosmos.
Selection of the correct treadmill device type according to the "list of
device types" in the appendix. Only the device type 0.1 and 1.1 can
be confirmed on the ladder ergometer.
Note: Selection and confirmation of a device type returns all user- and
administrator options to default settings.
For safety reasons, treadmill models with a motor hood or a crossbar
handrail must limit their speed to a maximum of 5 km/h during belt
rotation since they have an "obstacle" (motor hood or cross-bar
handrail) behind the subject during this mode.
Changing this speed limitation for these models is only permissible if
the safety of the runner is guaranteed (e.g. by a safety harness).
=
adjustable range from 1 km/h to maximum treadmill speed.
 selected unit is blinking,  max. is blinking
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
author: fh
optional settings: administrator options
indicates: maximum speed for reverse belt rotation,
email@h-p-cosmos.com
created 07.02.2013
§
with "donE".
printed 04.03.2013
page: 135 of 197