11.4 Running machines, oversize 200-450/75-300 (venus / saturn)

External control unit /
UserTerminal (display &
keyboard)
Power supply AC
(standard, read nameplate),
dedicated line required.
Fuse (standard, read nameplate)
Power input (long time)
Power input (momentary)
Drive motor capacity
Elevation motor capacity
Safety standards
Leakage current
Isolation transformer
Degree of protection
Overvoltage category
Mode of operation
Field of application / accuracy
according to EN 957
Classification according to MDD
Speed / definition
Speed accuracy*
Option special speed
Angle of inclination**
Definition
© 2013 h/p/cosmos sports & medical gmbh
(peak performances for milliseconds; max. 1s/min during heel strike)
MDD, directive 93/42/EEC + 2007/47/EC,
IEC EN 60601-1 (VDE 0750), IEC EN 60601-1-2, IEC EN 60601-1-4
VDE 0751, EN 957-1, EN 957-6, Machinery Directive 2006/42/EC
Category III: Life-Earth < 300 Volt = transient over-voltage limit 4000 Volt
continuous operation with intermittent loading (according to IEC 60601-1)
[r] = suitable for bicycle and wheelchair application (from running surface 3000 x 1250 mm standard)
active therapeutic device and active diagnostic device, risk class II b
0.0 ... 30.0 or ... 50.0 or ... 60.0 or ... 80.0 km/h
on request higher speed versions than 40 km/h only for sports models
-35 % ... +35 %
0.1 %
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
author: fh
email@h-p-cosmos.com
External control unit with TouchScreen
200 V 3N~ / 208 V 3N~ / 400 V 3N~
with N and PE
3 x C 32 A at 400 V 3N~
7500 VA
15200 VA
11000 VA
370 VA
n
0.25 mA
3 x 3000 VA
class I M / type B R / IP 20
sports and medical / S,I,A (EN957)
0.0 ... 40.0 km/h / 0.1 km/h
0.0 ... 11.11 m/s / 0.1 m/s
+/- 5 % (class A based on EN 957)
-27 % ... +27%**
0.1 %
created 07.02.2013
printed 04.03.2013
technical data
3 x 63 A
3 x 40 A
1 x 16 A
41500 VA
66500 VA
30000 VA
18000 VA
on request
3 x 21000
VA
-5 %
(-25 %) ...
+25 %