5.9.5

Test mode

Test mode provides several pre-defined tests. Eight tests can additionally be defined and stored by the user.
The first 11 tests represent standard ergometric tests. Furthermore, six training profiles, a walk protocol, profiles
VO2/10k, VO2/11k, VO2/12k, VO2/14k, profile Super Balke and test profiles 90-94 are available. They are
described in detail in the following chapters.
The UKK walk test (no. 1) is the only test where the running machine makes a calculation and an evaluation
(Fitness Index, but not for medical evaluation!). All other test profiles only control the load. There is no automatic
evaluation done by the running machine. The evaluation is to be done by host equipment (e.g. ECG,
ergospirometry, etc.) or external PC software.
Some test profiles (e.g. CONCONI test, Graded test, COOPER test, etc.) are endurance
N
tests (max. load and max. heart rate tests) and should only be performed after consulting a
medical doctor and under supervision of trained staff.
When using the treadmill for load tests, a medical doctor and a defibrillator must always be
N
within reach.
Proper warm-up and cool-down periods have to be considered.
N
Use a safety arch with chest belt system when making max. load and max. heart rate tests.
N
The following table shows the operating steps to run the treadmill via firmware or software h/p/cosmos para control
Operation of test mode
Initiation: running machine is in standby mode. LED mode is flashing.
Step no. Activity
[01]
Select test mode
[02]
Confirm test mode
[03]
Select test number
[04]
Confirm selected test
[05]
Stop test and running belt
© 2013 h/p/cosmos sports & medical gmbh
Keys
Response / Display
Selection of mode (manual, profile, cardio, test)
a
d
until
or
c
$
a
d
$
or
N
o
N
o
N
o
Display and reaction depend on the selected type of program /
c
test profile
Running belt stops.
b
(Deceleration time can be adjusted in the optional functions)
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
author: fh
email@h-p-cosmos.com

test is flashing
indicates program no. Pr. 1 is flashing
indicates program no. Pr. 1 ... "Pr. 28" is flashing
N
o
is illuminated
01 ... 11 predefined
12 ... 20 reserved
21 ... 28 user defined (free programmable profiles)
created 07.02.2013
printed 04.03.2013
operation
page: 53 of 197

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Venus 200/75450/300200/75170/65190/65

Table of Contents