5.4.7 Dual High Level Voltage Inputs
Refer to Figure 5-8 and wire the controller according to the procedure given in Table
5-10.
Table 5-10 Set Up Wiring Procedure for Dual High Level Voltage Inputs
Step
Connect the copper leads from the calibrator to the input to be calibrated as shown in
1
Figure 5-8.
2
Place voltage source at zero before switching on.
3
Following calibration, turn off the voltage source prior to disconnecting it from the
instrument.
Figure 5-8 Wiring Connections for Dual High Level Voltage Inputs
10/05
+
Millivolt or
Volt Source
_
Terminals for Input 1 are 35 (+) and 36 (-)
Terminals for Input 2 are 32 (+) and 33 (-)
Terminals for Input 3 are 29 (+) and 30 (-)
Terminals for Input 4 are 31 (+) and 33 (-)
Terminals for Input 5 are 28 (+) and 30 (-)
UDC3500 Universal Digital Controller Product Manual
Action
28+
Input 5
29+
Input 3
30-
31+
Input 4
32+
Input 2
33-
34R
35+
Input 1
36-
Input Calibration
269