Figure 1.16: Camera Calibration Menu; Figure 1.17: Camera Calibration Timer Menu; Figure 1.18: Feed Control Sheets Menu - Bell and Howell Spectrum XF Maintenance Manual

Scanner
Table of Contents

Advertisement

Note
For best results, the scanner should be warmed up for at least 40 minutes
before running camera calibration. Camera calibration requires that you feed
two clean calibration control sheets from the Camera Calibration Kit in
landscape operation.
Caution
Before calibrating either of the scanner's cameras, make sure to close all
open applications on the host PC before you begin. If an open application
attempts to communicate with the scanner during the calibration process,
irreversible problems with the scanner's EGSA board could result!
To calibrate the camera(s), perform the following:
1. Set the scanner in Manual Feeder Mode (refer to the Spectrum XF Operator Manual
for instructions).
2. From the Home menu, press the LEFT ARROW key until the Camera Calibration
menu is displayed.
3. Press one of the following FUNCTION keys: F1 [FRONT]; F2 [BACK]; or F3 [BOTH],
depending on which camera you wish to calibrate. The camera calibration icon will
indicate which camera needs to be calibrated.
4. During the first 40 minutes of scanner operation, the following menu will prompt you
to wait until the scanner and lamps are fully warmed up before calibrating.
5. Press the F1 key [OK] to cancel the calibration and return to the previous menu or
press the F3 key [CALIBRATE] to ignore the warm-up requirement and proceed with
the calibration.
Caution
Ignoring the warm-up requirement may result in degrading color image quality
and is recommended only after lamp replacement. If calibration is performed
during the warm-up period, the scanner will continue to prompt for re-
calibration until a successful calibration is performed after 40 minutes.
6. The control panel LCD prompts the operator to feed the control sheets.
S006504 Rev D
Camera Calibration
FRONT* BACK

Figure 1.16: Camera Calibration Menu

Wait
XX min for warm up?
OK
CALIBRATE

Figure 1.17: Camera Calibration Timer Menu

Feed white pages
CANCEL

Figure 1.18: Feed Control Sheets Menu

1 Operator Maintenance
BOTH
11 of 27