Dac Scaling; Control Register (W) Base + A H; Table B-4 Dac Scaling Examples - Agilent Technologies E6198B User Manual

Functional test system switch/load unit
Table of Contents

Advertisement

N O T E

DAC Scaling

Control Register (W) Base + A
E6198B Switch/Load Unit User Manual
To set the value of the DAC output, always write the MSB first, followed
by the LSB (register offset 9
the LSB is written. See
Bits
7
6
Write
DAC1
DAC1
7
6
The DAC output voltage is determined by the following:
V
= (N/16,384 32) - 16
out
where N = Decimal value of DAC code programmed; MSB, LSB.
Table B-4
shows some example values of N, and the
corresponding MSBs, LSBs, and DAC voltage outputs.

Table B-4 DAC Scaling Examples

N
Hex
16383
3FFF
10
12288
3000
10
8192
2000
10
4096
1000
10
0
0000
10
To reset the Switch/Load Unit including DACs, Open Drain
outputs and all Load and Pin Cards, write a 1 to this register,
wait 5 mS and then write a 0 to this register. DACs will reset to
0 Vout.
Bits
7
6
Write
x
x
). The output of the DAC will not update until
h
"DAC
Scaling".
5
4
3
DAC1
DAC1
DAC1
5
4
MSB
LSB
11111111
11111111
h
00110000
00000000
h
00100000
00000000
h
00010000
00000000
h
00000000
00000000
h
h
5
4
3
2
1
0
x
x
x
x
x
Reset
Register Definitions
2
1
DAC1
DAC1
DAC1
3
2
1
DAC V
out
+16V
+8V
0V
-8V
-16V
B
0
0
B-7