Switch/Load Unit And Card Specifications

Agilent U7177A 24-Channel Medium-Current Load Card Specifications

Parameter
Path resistance from power bus to load card
connect
Load path relay operate/release time
Peak Voltage to Earth
Load carry/switching current
Relay switching voltage
Minimum permisissible load
Basic current sense accuracy with standard 0.05
sense resistor
* Exclusive of load, with standard 0.05 sense resistor and 10A slo-blo fuse installed. (5 for load currents <0.1A.)
A-6
Specification
*
0.5
maximum
3.3/2.4 msec typical
8/8 msec max,
10 cps maximum
60 V continuous
500 V maximum transient, non-switching (delays to
<60V in 200 mS)
2 A maximum
60 Vdc maximum
10
10 mV DC
0.1% maximum