Agilent Technologies E6198B User Manual page 233

Functional test system switch/load unit
Table of Contents

Advertisement

E6198B Switch/Load Unit User Manual
Select
9
K17,
Relays
K18
Load Select 1-8 (W) Base + 4
This register controls the Load Select switch armature relays for
channels 1- 8. The register uses positive logic: 1 = closed.
Writing to this register starts the armature relay timer.
Bit
7
Purpose
K56
CH 8
Setting
state
• Write Only
• State: 1 = closed, 0 = open
• Power On/Reset State = 0
Load Select 9-16 (W) Base + 5
This register controls the Load Select switch armature relays for
channels 9-16. The register uses positive logic: 1 = closed.
Writing to this register starts the armature relay timer.
Bit
7
6
Purpose
K72
K70
CH 16
CH 15
Setting
state
state
• Write Only
• State: 1 = closed, 0 = open
• Power On/Reset State = 0
Power Select 1-8 (W) Base + 6
This register controls the Pull Up/Down Power Select armature
relays for channels 1-8. These relays are Form C and the register
uses positive logic: 1 = NO shorted to COM, NC is open; 0 = NC
10
11
12
K19,
K21,
K23,
K20
K22
K24
h
6
5
4
K54
K52
K50
CH 7
CH 6
CH 5
state
state
state
h
5
4
K68
K66
K64
CH 14
CH 13
CH 12
state
state
state
h
Register Definitions
13
14
15
K25,
K27,
K29,
K31,
K26
K28
K30
3
2
1
K48
K46
K44
K42
CH 4
CH 3
CH 2
CH 1
state
state
state
state
3
2
1
K62
K60
CH 11
CH 10
state
state
B
16
K32
0
0
K58
CH 9
state
B-17