Load And Uut Connections - Agilent Technologies E6198B User Manual

Functional test system switch/load unit
Table of Contents

Advertisement

5
Using Load Cards and Loads

Load and UUT Connections

5-60
Varistor voltage @ 1.0mA DC: 300 Vdc Max
Maximum Clamping Voltage (8/20mS): 450 volts
Capacitance: 17 pF
®
Typical Transzorb
specifications are:
Breakdown voltage: 300V max at 1mA
Reverse Standoff voltage: 250 volts
Maximum reverse leakage: 5mA
Maximum peak pulse current: 5A
Maximum Clamping voltage: 400 volts
Maximum temperature coefficient: 0.110%/°C
Figure 5-42
shows how the Agilent E6178B Load Card is wired.
Connect loads between J11-J18 and J31-J38. Loads are external
to the load card and connected to the load card via cables.
Connectors J11 and J31 connect to Channel 1, connectors J12
and J32 connect to Channel 2, etc.
When configured as part of a standard Agilent system, P2 of the
load card is typically connected via cables to a mass
interconnect panel. User connections to the UUT are then made
from the mass interconnect panel. Refer to the appropriate
mass interconnect wiring guide for connection details.
You can also make UUT and power supply connections directly
to P2 as shown in
Figure
5-42.