Protection Devices; Figure 5-46 N9377A Load Card - Flyback Circuit Detail - Agilent Technologies E6198B User Manual

Functional test system switch/load unit
Table of Contents

Advertisement

Figure 5-46 N9377A Load Card - Flyback Circuit Detail

UUT
loads
P2
Chan1

Protection Devices

E6198B Switch/Load Unit User Manual
In
Figure
input Form C switch. RT502 connects input line J4-J7, and
RT501 connects alternate input line J8-J11, to the output of
Load1. When a voltage spike occurs on the UUT that exceeds the
rating on the flyback device, the device clamps the surge voltage
to the device's predetermined value. The flyback protection is
installed similarly on each input line.
External
mount
loads
Load
switch
Load1.1
Load1.2
Load1
Install the flyback protection devices with the positive or
negative side towards the UUT as appropriate for your
application. On each of the 16 channels, note the location of the
polarization dot as shown in the flyback component locator
diagram,
back-to-back zener diodes are recommended for flyback voltage
protection.
Zener Diodes, MOVs (Metal-Oxide Varistor) or Transzorbs
devices mounted at RTx or RTy (2 required per load - 1 at NC
and 1 at NO) provide current path for the inductive load
flyback. Select the protection device so that it conducts at a
voltage higher than the UUT's internal protection. If the device's
internal protection fails, then the added external protection
conducts to protect the UUT and the load card.
5-46, Channel 1 has two input lines connected to the
ISense
-
1Pwr
LdSense-1
Figure
5-45. MOV (Metal Oxide Varistor), or
Using Load Cards and Loads
ISense
+
PB1
R
NO1
S
PwrSw1
NC1
LdSense+1
RT502
RT501
5
J4-J7
1
2
J8-J11
®
5-67

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ts-5000