Configuring the Switch/Load Unit

Configuring the Power Busses

4-6
The UUT power supplies attach to the power busses PB1-PB4.
Figure 4-4
shows the power bus connectors J801 and J802
located on the Switch/Load Unit backplane PC board for system
integrated SLU. These connectors use screw terminations for
high current capability. The J801 connectors are bussed
together on the PC board providing a common connection for up
to three supplies.

Figure 4-4 Power Bus Connectors J801 and J802

PB1
J801
The following examples A and B show the two most common
ways of configuring the power busses.
Example A shows three separate supplies attached to the
busses, with all three sharing a common ground on Power Bus 1
(PB1). This is the factory default configuration. The grounds for
all three supplies are connected together on connector J801.

Figure 4-5 Example A: Three Separate Supplies on PB1 - PB4

+12V
PC Board Traces
PB2
-12V
+5V
E6198B Switch/Load Unit User Manual
PB3
PB4
J802
PB1
PB2
PB3
PB4