E6198B Switch/Load Unit User Manual
shorted to COM, NO is open. The power buses selected are
determined by how the card is configured. Writing to this
register starts the relay timer.
7
6
5
Pwr8
Pwr7
Pwr6
state
state
state
• Write Only
• State: 1 = NO, shorted to Pwr node
• State: 0 = NC, shorted to Pwr node
• Power On/Reset State = 0 (all channels connected to NC)
Power Select 9-16 (W) Base + 13
This register controls the Pull Up/Down Power Select armature
relays for channels 9-16. These relays are Form C and the
register uses positive logic: 1 = NO shorted to COM, NC is open;
0 = NC shorted to COM, NO is open. The power buses selected
are determined by how the card is configured. Writing to this
register starts the relay timer.
7
6
5
Pwr16
Pwr15
Pwr14
state
state
state
• Write Only
• State: 1 = NO, shorted to Pwr node
• State: 0 = NC, shorted to Pwr node
• Power On/Reset State = 0 (all channels connected to NC)
Power Select 17-24 (W) Base + 14
This register controls the Pull Up/Down Power Select armature
relays for channels 17-24. These relays are Form C and the
register uses positive logic: 1 = NO shorted to COM, NC is open;
4
3
2
Pwr5
Pwr4
Pwr3
state
state
state
h
4
3
2
Pwr13
Pwr12
Pwr11
state
state
state
h
B
Register Definitions
1
0
Pwr2
Pwr1
state
state
1
0
Pwr10
Pwr9
state
state
B-45