Agilent E6175A 8-Channel High-Current Load Card Specifications - Agilent Technologies E6198B User Manual

Functional test system switch/load unit
Table of Contents

Advertisement

Agilent E6175A 8-Channel High-Current Load Card Specifications

Parameter
Path resistance from power bus to load card
connect
Load path relay operate/release time
Peak voltage to Earth
Load carry/switching current
Load carry peak current
Load switching power
Load switching voltage
Average power for loads mounted on load card
Minimum permissible load
Basic current sense accuracy with standard 0.05W
sense resistor
Maximum Current Consumption
E6198B Switch/Load Unit User Manual
Switch/Load Unit And Card Specifications
Specification
0.25
maximum
(Exclusive of load, with standard 0.05
and 10 A slo-blo fuse installed. (5
<1A.)
10/8 msec typical, 16/10 msec maximum, 30 cps
maximum
60 V continuous
500 V maximum transient, non-switching (delays to
<60V in 200 mS)
7.5 A maximum non-switching, continuous
15 A maximum, non-switching (<100 msec, <2% duty
cycle)
150 Watts maximum, resistive load
60 V maximum
40 Watts maximum total, per card
1 mA, 1V
0.1 % maximum
0.4 A at 5 V
0.3 A at 12 V
A
sense resistor
for load currents
A-3