Agilent Technologies E6198B User Manual page 230

Functional test system switch/load unit
Table of Contents

Advertisement

B
Register Definitions
B-14
Load Select (W) Base + 4
This register controls the Load Select switch armature relays of
the card, one per channel. The register uses positive logic: 1 =
closed. Writing to this register starts the armature relay timer.
Bit
7
6
Purpose
K43
K42
CH 8
CH 7
Setting
state
state
• Write Only
• State: 1 = closed, 0 = open
• Power On/Reset State = 0
Power Select (W) Base + 5
This register controls the Pull Up/Down Power Select armature
relays of the card, one per channel pair. These relays are Form
C (double-pole, single-throw) and the register uses positive
logic: 1 = Normally open (NO) shorted to COM, normally closed
(NC) is open; 0 = NC shorted to COM, NO is open. The power
buses selected depend on how the card is configured by the
user. Writing to this register starts the armature relay timer.
Bit
7-4
Purpose
Undefined
Setting
X
• Write Only
• State: 1 = NO shorted to COM, NC is open; 0 = NC shorted to
COM, NO is open
• Power On/Reset State = 0
h
5
4
3
K33
K32
K23
CH 6
CH 5
CH 4
state
state
state
h
3 (011)
2 (010)
1 (001)
K41
K31
K21
CH 8/7
CH 6/5
CH 4/3
state
state
state
E6198B Switch/Load Unit User Manual
2
1
0
K22
K13
K12
CH 3
CH 2
CH 1
state
state
state
0 (000)
K11
CH 2/1
state

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ts-5000

Table of Contents