Agilent Technologies E6198B User Manual page 240

Functional test system switch/load unit
Table of Contents

Advertisement

B
Register Definitions
B-24
controlling the respective relays are written to or the card is
reset. The card remains in the busy state until both timers have
timed out.
Bit
7-1
Purpose
Undefined
Setting
all is 1s
• Read Only
• Undefined bits readback as all 1s
• BUSY~: 0 = Busy, 1 = Ready
Current Sense Select (W) Base + 3h
This register controls the current sense relays of the card. Since
only one current sense channel at a time per slot is allowed,
these channel selects are encoded. This allows for none or one
current sense relay to be selected. Writing to this register starts
the reed relay timer.
Bit
7-6
Purpose
Undefined
Setting
• Write Only
• Undefined bits are not used
• Valid current sense selection values are: 000001
Selecting values 000000
select current sensing for any channel.
000001
- corresponds to current sense select on channel 1.
2
000010
- corresponds to current sense select on channel 2.
2
001101
- corresponds to current sense select on channel 13.
2
010000
- corresponds to current sense select on channel 16.
2
000000
or nnnnnn
2
selected.
0
BUSY~
state
5-0
Sense Select
X
Select No.
or 011001
- 111111
2
2
> 011000
- no current sense relays
2
2
E6198B Switch/Load Unit User Manual
- 011000
.
2
2
will NOT