Instrument Setup; Recalling A State File - Agilent Technologies E5071C Compliance Tests

Mhl (mobile high-definition link) specification version 2.0; agilent method of implementation (moi) for mhl cables; ena option tdr
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Agilent MOI for MHL Cables Compliance Tests

5.2. Instrument Setup

5.2.1. Recalling a State File

This section describes how to recall a state file of the E5071C that includes all the
measurement settings for MHL cable compliance test. The state file can be downloaded at:
www.agilent.com/find/ena-tdr_mhl-cabcon
Copy the state file into the E5071C's directory via USB mass storage device and recall the
state file using the TDR software. Necessary parameters for testing are automatically set up
in the E5071C. Refer to Appendix for the details about manual setup.
If TDR setup wizard is shown, click
1. Open
Setup
tab.
2. Click
Advanced Mode
3. A dialog box appears requesting for confirmation. Then click Yes. (Uncheck "Use
Advanced Calibration Methods")
4. Click
File
and select
5. Specify a folder and a file name, and click Open.
The E5071C's channel 1 is used for time-domain measurements by using the TDR software
at the bottom of the E5071C's screen. The channel 2 is used for frequency-domain
Close
button in the TDR setup wizard main window.
to show the dialog box.
Recall State
to open the Recall State dialog box.