Agilent MOI for MHL Cables Compliance Tests
5. Select Format to "Volt".
6. Click
Marker Search
7. Press
Run
to measure the rise time on the screen.
8. Enter rise time until the measure rise time is close to the specified value.
 "200 ps" for Differential Characteristic Impedance.
 "600 ps" for Common-Mode Characteristic Impedance.
9. Click
Marker Search
10. Select Format to "Impedance".

5.3.2. Frequency Domain Calibration

The purpose of this step is to calibrate out the RF effects (i.e. mismatch, loss or delay) of
RF cables and test fixtures. Full calibration is performed by using the 4-port ECal Module
at the end of RF cables connected to the ENA test ports. The fixture effect will be
eliminated by de-embedding S-parameters (*.s4p or .s2p) of the fixture. The S-parameter
data is provided by fixture vendors.
The calibration is applied for frequency-domain measurements in channel 2.
5.3.2.1.
Calibration for the frequency-domain measurement is performed by selecting the ENA's
soft key. The 4-port ECal Module (i.e. N4431B) connected to the E5071C is necessary for
the calibration procedure.
1. Press
Channel Next
and select "Rise Time (20 – 80%)".
and select "Rise Time (20 – 80%)" to turn off the marker.
ECal Calibration
key to select Channel 2.
12