Common-Mode Characteristic Impedance - Agilent Technologies E5071C Compliance Tests

Mhl (mobile high-definition link) specification version 2.0; agilent method of implementation (moi) for mhl cables; ena option tdr
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Agilent MOI for MHL Cables Compliance Tests
12. Click
Trace Settings Copy
13. Select the Trace 1 in the From list.
14. Select the Trace 7 in the To list.
15. Click Copy.
16. Click Close.
17. Select
Trace
7.
18. Open
Parameter
19. Click Tdd22.

6.5. Common-Mode Characteristic Impedance

1. Select
Trace
2.
2. Open
Parameters
4. Select "Time Domain' and "Differential" for Measure.
5. Select Format to "Impedance"
6. Click Tcc11.
7. Click the box below the left knob under Horizontal. Set the horizontal scale to "400
ps/div" in a dialog box.
8. Click the box below the right knob under Horizontal. Set the horizontal center to "2 ns"
in a dialog box.
9. Click the box below the left knob under Vertical. Set the vertical scale to "2 Ohm/div"
in a dialog box.
10. Click the box below the right knob under Vertical. Set the vertical center to "20 Ohm"
in a dialog box.
11. Open
Trace Control
12. Click
Trace Settings Copy
13. Select the Trace 2 in the From list.
14. Select the Trace 8 in the To list.
15. Click Copy.
to launch trace copy dialog box.
tab.
tab.
tab.
to launch trace copy dialog box.
31