Agilent MOI for MHL Cables Compliance Tests

6.3.2. Crosstalk Compensation

1. Select
Trace 4.
> Equation Editor... > Enter an equation "S31+S32".
2. Press
Display
3. Check Enabled to enable the equation on trace.
4. Click Apply.
5. Click Close.
6. Select
Trace 10.
> Equation Editor... > Enter an equation "S42+S41".
7. Press
Display
8. Check Enabled to enable the equation on trace.
9. Click Apply.
10. Click Close.

6.4. Differential Characteristic Impedance

1. Select
Trace
1.
2. Open
Parameters
4. Select "Time Domain' and "Differential" for Measure.
5. Select Format to "Impedance"
6. Click Tdd11.
7. Click the box below the left knob under Horizontal. Set the horizontal scale to "200
ps/div" in a dialog box.
8. Click the box below the right knob under Horizontal. Set the horizontal center to "1 ns"
in a dialog box.
9. Click the box below the left knob under Vertical. Set the vertical scale to "10 Ohm/div"
in a dialog box.
10. Click the box below the right knob under Vertical. Set the vertical center to "50 Ohm"
in a dialog box.
11. Open
Trace Control
tab.
tab.