Agilent MOI for MHL Cables Compliance Tests
2. Connect all test cables to the ECal Module.
3. Press
Calibrate
> ECal > 4-Port Cal.
5.3.2.2.
The effect of the test fixtures are removed by a de-embedding function of the E5071C's
firmware. 4-port Touchstone files (*.s4p) are used to compensate differential transmission
lines of the fixtures, while 2-port Touchstone files (*.s2p) are used for compensation of the
single-ended lines.
The following 4-port fixture compensation is applied to 4-port measurements such as
Differential
Insertion
Common-Mode conversion, and Far-End Crosstalk measurements.
1. Press
Trace Next
2. Press
Analysis
> Fixture Simulator and turn it ON.
3. Press
Analysis
> Fixture Simulator > De-Embedding SnP > Topology > Select
Topology to select topology C.
4. Press
Analysis
> Fixture Simulator > De-Embedding SnP > Topology > Ports > to
select 1-2-3-4.
5. Press
Analysis
> Fixture Simulator > De-Embedding SnP > Topology > User File
(nwk1) and specify a 4-port de-embedding file (*.s4p) for the Source fixture.
6. Press
Analysis
> Fixture Simulator > De-Embedding SnP > Topology > Type
(nwk1) and select De-embed.
7. Press
Analysis
> Fixture Simulator > De-Embedding SnP > Topology > User File
(nwk2) and specify a de-embedding file (*.s4p) for the Sink fixture.
8. Press
Analysis
> Fixture Simulator > De-Embedding SnP > Topology > Type
(nwk2) and select De-embed.
De-embedding Fixture
Loss,
Common-Mode
to select trace.
Insertion
13
Loss,
Differential
and