Agilent Technologies E5071C Compliance Tests page 24

Mhl (mobile high-definition link) specification version 2.0; agilent method of implementation (moi) for mhl cables; ena option tdr
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Agilent MOI for MHL Cables Compliance Tests
5.4.11.2. Measurement and Data Analysis
1. Connect DUT (Figure 5-4).
2. Press
Trace Next
3. Press
Start
> and set start frequency to 1 MHz.
4. Press
Stop
> and set stop frequency to 5.1 GHz.
5. Press
Trigger
> Single.
6. Confirm the measured far-end crosstalk is lower than the limit shown in Table 5-9.
7. Press
Trace Next
8. Confirm the measured far-end crosstalk is lower than the limit shown in Table 5-9.
Table 5-9 Far-End Crosstalk Limit between MHL and CBUS or VBUS
Frequency
>30 M to 300 MHz
>300 M to 5.1 GHz
(*) Excursions are permitted to -11.5 dB for the frequency range of 825 to 5100 MHz if
the differential insertion loss of the cable meets the below:
Table 5-10 Differential Insertion Loss Limit for Far-End Crosstalk Excursion
Frequency (MHz)
9. Repeat the same measurements for all the connections in Table 5-8. When measuring
crosstalk between CBUS and VBUS lanes, confirm the measured crosstalk is lower
to select Trace 3 (S41).
to select Trace 7 (S32).
0 to 30 MHz
0
825
2475
4125
5100
Attenuation (dB)
-30
-20
-15 (*)
Attenuation (dB)
2.0
2.0
8.0
14.5
18.0
24