Agilent MOI for MHL Cables Compliance Tests
11. Input
Position (%)
12. Click OK.
13. Confirm the measured intra-pair skew is within the limit. If common-mode intra-pair
skew is less than or equal to 43 ps, then PASS. Otherwise FAIL.

5.4.2. Common-mode Intra-Pair Skew

1. Connect DUT (Figure 5-1).
2. Select
Trace
4.
3. Press
Stop
Single.
4. Click
Auto Scale
transition point of x-axis (in sec).
5. Click the box below the right knob under Horizontal. Set the horizontal value to the
above transition point in a dialog box, so that the transition point can be centered.
6. Select
Trace 10
and repeat the same operations, step 4 to 5.
7. Select
Trace
4.
8. Click
Marker Search
Δ
9. Check the
Time.
10. Select
Target (Stop)
11. Input
Position (%)
12. Click OK.
13. Confirm the measured intra-pair skew is within the limit. If common-mode intra-pair
skew is less than or equal to 43 ps, then PASS. Otherwise FAIL.
to 50.
and select "X & Y" for auto scaling the x-axis and y-axis. Record the
Δ Time
and select
to Trace 10 (T42).
to 50.
to show the dialog box.
16