Agilent MOI for MHL Cables Compliance Tests
8. Press
Analysis
> Fixture Simulator > Topology > Port2 (bal) > 3-4
9. Press Display.> Num of Traces > 8.
10. Press Display.> Allocate Traces > x4 (2 columns by 2 rows).
6.8. Differential Insertion Loss
1. Press
Trace Next
2. Press
Analysis
> Fixture Simulator > BalUn and turn it ON
3. Press
Analysis
> Fixture Simulator > Measurement > Sdd21.
4. Set Scale/Div to 5 dB/div.
5. Set Reference Value to -10 dB.
6.9. Common-Mode Insertion Loss
1. Press
Trace Next
2. Press
Analysis
> Fixture Simulator > BalUn and turn it ON
3. Press
Analysis
> Fixture Simulator > Measurement > Scc21.
4. Set Scale/Div to 2 dB/div.
5. Set Reference Value to -10 dB.
6.10. Differential and Common-mode Conversion
1. Press
Trace Next
2. Press
Analysis
> Fixture Simulator > BalUn and turn it ON
3. Press
Analysis
> Fixture Simulator > Measurement > Sdc21.
4. Set Scale/Div to 5 dB/div.
5. Set Reference Value to -20 dB.
6. Press
Trace Next
to select Trace 1.
to select Trace 5.
to select Trace 2.
to select Trace 6.
33