Cbus Cable Delay - Agilent Technologies E5071C Compliance Tests

Mhl (mobile high-definition link) specification version 2.0; agilent method of implementation (moi) for mhl cables; ena option tdr
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Agilent MOI for MHL Cables Compliance Tests
Table 5-2 Common-mode Characteristic Impedance Limit
MHL Clock Common-mode Impedance
*1. Connection point and transition area, up to 3 ns distance. A single excursion is
permitted out to 20-40 ohms, no wider than 500 ps.
*2. Cable area, from 3 ns to 4 ns distance.

5.4.5. CBUS Cable Delay

1. Connect DUT (Figure 5-2).
Figure 5-2. CBUS Cable Delay Test Setup
2. Select
Trace
5.
3. Press
Stop
Single.
4. Click
Auto Scale
and select "X" for auto scaling the x-axis.
Min
Max
Unit
24
36
Ohm
25
35
Ohm
18
Note
*1
*2