Agilent MOI for MHL Cables Compliance Tests
9. Click
Measure
mark appears, click Finish.
10. Read DUT Length and confirm it is over 18 ns. Otherwise input 18 ns in the field.
5.3.1.2.
The effective rise time is adjusted for the specification of characteristic impedance
measurements. DUT is disconnected during the adjustment procedure.
1. Open
TDR/TDT
2. Click
Trace
x.
 Trace 1 & Trace 7 for Differential Characteristic Impedance.
 Trace 2 & Trace 8 for Common-Mode Characteristic Impedance.
3. Click
Parameter
4. Select "Time Domain" and "Differential" for Measure.
to specify DUT's electrical length in the dialog box. Once green check
Adjustment of Effective Rise Time
tab.
tab.
11