Agilent MOI for MHL Cables Compliance Tests
6. Select
Trace 9.
> Equation Editor... > Enter an equation "-S42+S41".
7. Press
Display
8. Check Enabled to enable the equation on trace.
9. Click Apply.
10. Click Close.

6.3. Common-Mode Intra-Pair Skew

6.3.1. Parameter Setup

1. Open
Trace Control
2. Click
Trace Settings Copy
3. Select the
Trace 3
4. Select the
Trace 4
5. Select the
Trace 10
6. Click Copy.
7. Click
Close
8. Select
Trace
10.
9. Open
Parameters
10. Click T42.
tab.
to launch trace copy dialog box.
in the From list.
in the To list.
in the To list.
tab.
29