Agilent MOI for MHL Cables Compliance Tests

5.4.11. Far-End Crosstalk

5.4.11.1. Connections
Figure 5-4 Far-End Crosstalk Test Setup (MHL- & CBUS)
The 4-port S-parameter measurements for Far-End Crosstalk are performed for the
following combinations (Table 5-7). The test ports of the E5071C (Port 1 to 4) are
connected to the ports of test fixtures.
Table 5-8 Test Port Connections for Far-End Crosstalk Measurements
Lines / Ports
MHL- & CBUS
MHL+ & VBUS
MHL- & CBUS
CBUS & VBUS
Port 1
Source MHL-
Source CBUS
Source MHL+ Source VBUS
Source MHL-
Source CBUS
Source CBUS
Source VBUS
23
Port 2
Port 3
Sink MHL-
Sink MHL+
Sink MHL-
Sink CBUS
Port 4
Sink CBUS
Sink VBUS
Sink CBUS
Sink VBUS