Differential Insertion Loss; Common-Mode Insertion Loss - Agilent Technologies E5071C Compliance Tests

Mhl (mobile high-definition link) specification version 2.0; agilent method of implementation (moi) for mhl cables; ena option tdr
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Agilent MOI for MHL Cables Compliance Tests
5. Press
Marker Search
6. Confirm the measured CBUS Cable Delay is within the limit. If the delay is less than or
equal to 35 ns, then PASS. Otherwise FAIL.

5.4.6. Differential Insertion Loss

1. Connect DUT (Figure 5-1).
2. Press
Channel Next
3. Press
Trace Next
4. Press
Start
> and set start frequency to 1 MHz.
5. Press
Stop
> and set stop frequency to 5.1 GHz.
6. Press
Trigger
> Single.
7. Confirm the measured differential insertion loss is lower than the limit shown in Table
5-3.
Table 5-3 Differential Insertion Loss Limit
Frequency (MHz)

5.4.7. Common-mode Insertion Loss

1. Connect DUT (Figure 5-1)
2. Press
Trace Next
3. Press
Start
> and set start frequency to 1 MHz.
> Target > Target Value and enter "0.1".
to select Channel 2.
to select Trace 1 (Sdd21).
0
825
2475
4125
5100
to select Trace 4 (Scc21).
Attenuation (dB)
2.25
5.5
12.0
18.0
21.5